How to Perform Measurements in the I/Q Analyzer Application
R&S
®
ZNL
95
User Manual 1178.5989.02 ─ 06
7
How to Perform Measurements in the I/Q
Analyzer Application
The following step-by-step instructions demonstrate how to capture I/Q data on the
R&S
ZNL and how to analyze data in the I/Q Analyzer application.
How to Capture Baseband (I/Q) Data as RF Input
..................................................95
How to Analyze Data in the I/Q Analyzer
................................................................96
7.1
How to Capture Baseband (I/Q) Data as RF Input
By default, the I/Q Analyzer assumes the I/Q data is modulated on a carrier frequency
and input via the "RF Input" connector on the R&S
ZNL.
1. Select the [MODE] key and select the "I/Q Analyzer" application.
2. Select the "Overview" softkey to display the "Overview" for an I/Q Analyzer mea-
surement.
3. Select the "Input" button to select and configure the "RF Input" signal source.
4. Select the "Amplitude" button to define the attenuation, reference level or other set-
tings that affect the input signal's amplitude and scaling.
5. Select the "Frequency" button to define the input signal's center frequency.
6. Optionally, select the "Trigger" button and define a trigger for data acquisition, for
example an I/Q Power trigger to start capturing data only when a specific power is
exceeded.
7. Select the "Bandwidth" button and define the bandwidth parameters for data acqui-
sition:
● "Sample Rate" or "Analysis Bandwidth" the span of the input signal to be cap-
tured for analysis, or the rate at which samples are captured (both values are
correlated)
● "Measurement Time" how long the data is to be captured
● "Record Length" : the number of samples to be captured (also defined by sam-
ple rate and measurement time)
8. Select the "Display Config" button and select up to six displays that are of interest
to you.
Arrange them on the display to suit your preferences.
9. Exit the SmartGrid mode.
10. Start a new sweep with the defined settings.
a) Select the Sequencer icon ( ) from the toolbar.
b) Set the Sequencer state to "Off".
How to Capture Baseband (I/Q) Data as RF Input