Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
126
User Manual 1178.5989.02 ─ 06
9.4.3
Frequency
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
................................................................ 126
......................................................................................... 126
................................................................................126
[SENSe:]FREQuency:CENTer:STEP:AUTO
......................................................................127
......................................................................................... 127
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
<m>
Example:
CALC:MARK2:FUNC:CENT
Sets the center frequency to the frequency of marker 2.
Manual operation:
See
" Center Frequency = Marker Frequency "
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency.
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP
AND
SENS:FREQ DOWN
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz