Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
107
User Manual 1178.5989.02 ─ 06
Example:
INST:LIST?
Result for 3 channel setups:
'ADEM','Analog Demod','IQ','IQ
Analyzer','IQ','IQ Analyzer2'
Usage:
Query only
Table 9-2: Available channel setup types and default channel setup names
Application
<ChannelType> Parameter
Default Channel Setup Name*)
Spectrum
SANALYZER
Spectrum
Analog Demodulation
ADEM
Analog Demod
I/Q Analyzer
IQ
IQ Analyzer
Noise Figure Measure-
ments
NOISE
Noise
Vector Signal Analysis
(VSA)
DDEM
VSA
Network analysis
Vna
VNA
Note: the default channel setup name is also listed in the table. If the specified name for a new channel
setup already exists, the default name, extended by a sequential number, is used for the new channel
setup.
INSTrument:REName
<ChannelName1>, <ChannelName2>
This command renames a channel setup.
Setting parameters:
<ChannelName1>
String containing the name of the channel setup you want to
rename.
<ChannelName2>
String containing the new channel setup name.
Note that you cannot assign an existing channel setup name to
a new channel setup; this will cause an error.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Usage:
Setting only
INSTrument[:SELect]
<ChannelType> | <ChannelName>
This command activates a new channel setup with the defined channel setup type, or
selects an existing channel setup with the specified name.
Also see
●
Activating I/Q Analyzer Measurements