![Rohde & Schwarz ZNL Series User Manual Download Page 199](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893199.webp)
Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
199
User Manual 1178.5989.02 ─ 06
Example:
CALC:MARK4:FUNC:BPOW:STAT ON
Activates or turns marker 4 into a band power marker.
Manual operation:
See
" Band Power Measurement State "
See
" Switching All Band Power Measurements Off "
Using Delta Markers
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
.............................................199
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
...........................................199
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN
..............................................200
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
........................................... 200
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
<Mode>
This command selects the way the results for a band power delta marker are dis-
played.
Suffix:
<n>
.
<m>
Parameters:
<Mode>
POWer
Result is displayed as an absolute power. The power unit
depends on the
DENSity
Result is displayed as a density in dBm/Hz.
RPOWer
This setting is only available for a delta band power marker.
The result is the difference between the absolute power in the
band around the delta marker and the absolute power for the
reference marker. The powers are subtracted logarithmically, so
the result is a dB value.
[Relative band power (Delta2) in dB] = [absolute band power
(Delta2) in dBm] - [absolute (band) power of reference marker in
dBm]
*RST:
POWer
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
<m>
I/Q Analysis