Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
108
User Manual 1178.5989.02 ─ 06
Parameters:
<ChannelType>
Channel type of the new channel setup.
For a list of available channel setup types see
<ChannelName>
String containing the name of the channel setup.
Example:
INST IQ
INST 'MyIQSpectrum'
Selects the channel setup named 'MyIQSpectrum' (for example
before executing further commands for that channel setup).
SYSTem:PRESet:CHANnel[:EXEC]
This command restores the default instrument settings in the current channel setup.
Use
INST:SEL
to select the channel setup.
Example:
INST:SEL 'Spectrum2'
Selects the channel setup for "Spectrum2".
SYST:PRES:CHAN:EXEC
Restores the factory default settings to the "Spectrum2" channel
setup.
Usage:
Event
Manual operation:
See
TRACe:IQ:EVAL
<State>
This command turns I/Q data analysis on and off.
Before you can use this command, you have to turn on the I/Q data acquisition using
INST:CRE:NEW IQ
or
INST:CRE:REPL
, or using the
mand to replace the current channel setup while retaining the settings.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
TRAC:IQ ON
Enables I/Q data acquisition
TRAC:IQ:EVAL ON
Enables the I/Q data analysis mode.
TRACe:IQ[:STATe]
<State>
Executing this command also has the following effects:
●
The sweep, amplitude, input and trigger settings from the previous application are
retained
Activating I/Q Analyzer Measurements