![Rohde & Schwarz ZNL Series User Manual Download Page 145](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893145.webp)
Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
145
User Manual 1178.5989.02 ─ 06
Table 9-3: <WindowType> parameter values for IQ Analyzer application
Parameter value
Window type
FREQ
Spectrum
MAGN
Magnitude
MTABle
Marker table
PEAKlist
Marker peak list
RIMAG
Real/Imag (I/Q)
VECT
I/Q Vector
LAYout:CATalog[:WINDow]?
This command queries the name and index of all active windows in the active channel
setup from top left to bottom right. The result is a comma-separated list of values for
each window, with the syntax:
<WindowName_1>,<WindowIndex_1>..<WindowName_n>,<WindowIndex_n>
Return values:
<WindowName>
string
Name of the window.
In the default state, the name of the window is its index.
<WindowIndex>
numeric value
Index of the window.
Example:
LAY:CAT?
Result:
'2',2,'1',1
Two windows are displayed, named '2' (at the top or left), and '1'
(at the bottom or right).
Usage:
Query only
LAYout:IDENtify[:WINDow]?
<WindowName>
This command queries the
index
of a particular display window in the active channel
setup.
Note
: to query the
name
of a particular window, use the
query.
Query parameters:
<WindowName>
String containing the name of a window.
Return values:
<WindowIndex>
Index number of the window.
Example:
LAY:WIND:IDEN? '2'
Queries the index of the result display named '2'.
Response:
2