Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
110
User Manual 1178.5989.02 ─ 06
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
ZNL after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
See
INPut<ip>:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
ZNL.
Suffix:
<ip>
.
1..n
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
*RST:
RF
Manual operation:
See
See
Configuring I/Q Analyzer Measurements