Analysis
R&S
®
ZNL
62
User Manual 1178.5989.02 ─ 06
"View"
The current contents of the trace memory are frozen and displayed.
"Blank"
Removes the selected trace from the display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Detector
Defines the trace detector to be used for trace analysis.
The trace detector is used to combine multiple FFT window results to create the final
spectrum. (Note: in previous versions of the R&S
ZNL, the I/Q Analyzer always used
the linear average detector.) If necessary, the trace detector is also used to reduce the
number of calculated frequency points (defined by the FFT length) to the defined num-
ber of sweep points. By default, the Autopeak trace detector is used.
Note:
Using a detector other than Auto Peak and fewer than 4096 sweep points may
lead to wrong level results. For details see
"Combining results - trace detector"
"Auto"
Selects the optimum detector for the selected trace and filter mode.
This is the default setting.
"Type"
Defines the selected detector type.
Remote command:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
Hold
If activated, traces in "Min Hold", "Max Hold" and "Average" mode are not reset after
specific parameter changes have been made.
Normally, the measurement is started again after parameter changes, before the mea-
surement results are analyzed (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
The default setting is off.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
Average Mode
Defines the mode with which the trace is averaged over several sweeps.
This setting is generally applicable if trace mode "Average" is selected. For FFT
sweeps, the setting also affects the VBW (regardless of whether or not the trace is
averaged).
(See the chapter on ACLR power measurements in the R&S
ZNL User Manual.)
How many sweeps are averaged is defined by the
Trace Settings