Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
160
User Manual 1178.5989.02 ─ 06
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
Suffix:
<n>
.
<t>
Parameters:
<Detector>
APEak
Autopeak
NEGative
Negative peak
POSitive
Positive peak
SAMPle
First value detected per trace point
RMS
RMS value
AVERage
Average
*RST:
APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
I/Q Analysis