Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
174
User Manual 1178.5989.02 ─ 06
Example:
CALC:MARK:AOFF
Switches off all markers.
Manual operation:
See
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
<State>
This command links normal marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Suffix:
<n>
.
<ms>
source marker, see
<md>
destination marker, see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK4:LINK:TO:MARK2 ON
Links marker 4 to marker 2.
Manual operation:
See
CALCulate<n>:MARKer<m>[:STATe]
<State>
This command turns markers on and off. If the corresponding marker number is cur-
rently active as a delta marker, it is turned into a normal marker.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation:
See
See
See
I/Q Analysis