![Rohde & Schwarz ZNL Series User Manual Download Page 53](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893053.webp)
Configuration
R&S
®
ZNL
53
User Manual 1178.5989.02 ─ 06
On
I and Q signals are interchanged
Inverted sideband, Q+j*I
Off
I and Q signals are not interchanged
Normal sideband, I+j*Q
Remote command:
RBW
Defines the resolution bandwidth for Spectrum results. The available RBW values
depend on the sample rate and record length.
(See
Chapter 4.2.4, "Frequency Resolution of FFT Results - RBW"
Depending on the selected RBW mode, the value is either determined automatically or
can be defined manually. As soon as you enter a value in the input field, the RBW
mode is changed to "Manual".
If the "Advanced Fourier Transformation Params" option is enabled, advanced FFT
mode is selected and the RBW cannot be defined directly.
Note that the RBW is correlated with the
). Changing any one of these parameters
may cause a change to one or more of the other parameters. For more information see
"Auto mode"
(Default) The RBW is determined automatically depending on the
"Manual mode" The RBW can be defined by the user.
The user-defined RBW is used and the
(and possibly
"Advanced
FFT mode"
This mode is used if the "Advanced Fourier Transformation Params"
option is enabled.
The RBW is determined by the
.
Remote command:
Advanced FFT mode / Basic Settings
Shows or hides the "Advanced Fourier Transformation" parameters in the "Data Acqui-
sition" dialog box.
Note that if the advanced FFT mode is used, the
settings are not available.
TransformationAlgorithm ← Advanced FFT mode / Basic Settings
Defines the FFT calculation method.
"Single"
One FFT is calculated for the entire record length; if the
is larger than the record length, zeros are appended to the captured
data.
Data Acquisition and Bandwidth Settings