Configuration
R&S
®
ZNL
52
User Manual 1178.5989.02 ─ 06
Advanced FFT mode / Basic Settings
..........................................................................53
................................................................................53
........................................................................................... 54
.............................................................................................54
.............................................................................................. 54
Sample Rate
Defines the I/Q data sample rate of the R&S
ZNL. This value is dependent on the
and the defined signal source.
sample rate = analysis bandwidth / 0.8
For details on the dependencies see
Chapter 4.1.1, "Sample Rate and Maximum Usa-
ble I/Q Bandwidth for RF Input"
Remote command:
Analysis Bandwidth
Defines the flat, usable bandwidth of the final I/Q data. This value is dependent on the
defined
and the defined signal source.
analysis bandwidth = 0.8 * sample rate
Remote command:
Meas Time
Defines the I/Q acquisition time. By default, the measurement time is calculated as the
number of I/Q samples ( "Record Length" ) divided by the sample rate. If you change
the measurement time, the
is automatically changed, as well.
Remote command:
Record Length
Defines the number of I/Q samples to record. By default, the number of sweep points is
used. The record length is calculated as the measurement time multiplied by the sam-
ple rate. If you change the record length, the
is automatically changed, as
well.
Note:
For the I/Q vector result display, the number of I/Q samples to record ( "Record
Length" ) must be identical to the number of trace points to be displayed ("Sweep
Points"). Thus, the sweep points are not editable for this result display. If the "Record
Length" is edited, the sweep points are adapted automatically.
Remote command:
Swap I/Q
Activates or deactivates the inverted I/Q modulation. If the I and Q parts of the signal
from the DUT are interchanged, the R&S
ZNL can do the same to compensate for it.
Data Acquisition and Bandwidth Settings