Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
155
User Manual 1178.5989.02 ─ 06
Note that the number of sweep points is limited to 10001 when measuring spurious
emissions.
Suffix:
<n>
.
Parameters:
<SweepPoints>
Range:
101 to 100001
*RST:
1001
Example:
SWE:POIN 251
Manual operation:
See
[SENSe<n>:]SWEep:TIME
<Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
Suffix:
<n>
.
1..n
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Example:
SWE:TIME 10s
Manual operation:
See
SYSTem:SEQuencer
<State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (
INIT:SEQ...
) are executed, otherwise an error will
occur.
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S
ZNL User Manual.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands
(
INIT:SEQ...
) are not available.
*RST:
0
Capturing Data and Performing Sweeps