Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
129
User Manual 1178.5989.02 ─ 06
Parameters:
<Period>
Range:
0 s to 10 s
*RST:
0 s
Default unit: S
Example:
TRIG:SOUR EXT
Sets an external trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation:
See
TRIGger[:SEQuence]:IFPower:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis, which is only available for "IF Power" trig-
ger sources.
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
*RST:
3 dB
Default unit: DB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal<port>]
<TriggerLevel>
This command defines the level the external signal must exceed to cause a trigger
event.
In the I/Q Analyzer application, only
EXTernal1
is supported.
Suffix:
<tp>
.
irrelevant
Parameters:
<TriggerLevel>
For the R&S
ZNL, the external trigger level is always 1.4
V. It
cannot be changed.
*RST:
1.4 V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation is considered when the trigger level is analyzed. If
defined, a reference level offset is also considered.
Configuring I/Q Analyzer Measurements