Configuration
R&S
®
ZNL
49
User Manual 1178.5989.02 ─ 06
Remote command:
TRIG:SOUR IMM
, see
ExternalTrigger 1 ← Trigger Source ← Trigger Source
Data acquisition starts when the TTL signal fed into the trigger input connector of the
R&S
ZNL meets or exceeds the specified trigger level.
(See
Remote command:
TRIG:SOUR EXT
See
IF Power ← Trigger Source ← Trigger Source
The R&S
ZNL starts capturing data as soon as the trigger level is exceeded around the
third intermediate frequency.
For frequency sweeps, the third IF represents the start frequency. The trigger band-
width at the third IF depends on the RBW and sweep type.
For measurements on a fixed frequency (e.g. zero span or I/Q measurements), the
third IF represents the center frequency.
This trigger source is only available for RF input.
For details on available trigger levels and trigger bandwidths, see the data sheet.
Remote command:
TRIG:SOUR IFP
, see
I/Q Power ← Trigger Source ← Trigger Source
This trigger source is only available in the I/Q Analyzer application and in applications
that process I/Q data.
Triggers the measurement when the magnitude of the sampled I/Q data exceeds the
trigger threshold.
The trigger bandwidth corresponds to the bandwidth setting for I/Q data acquisition.
Remote command:
TRIG:SOUR IQP
, see
Time ← Trigger Source ← Trigger Source
Triggers in a specified repetition interval.
Remote command:
TRIG:SOUR TIME
, see
Trigger Level ← Trigger Source
Defines the trigger level for the specified trigger source.
For details on supported trigger levels, see the data sheet.
Remote command:
TRIGger[:SEQuence]:LEVel:IFPower
TRIGger[:SEQuence]:LEVel:IQPower
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal<port>]
Trigger Settings