![Rohde & Schwarz ZNL Series User Manual Download Page 41](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893041.webp)
Configuration
R&S
®
ZNL
41
User Manual 1178.5989.02 ─ 06
Noise Source Control
The R&S
ZNL provides a connector ("NOISE SOURCE CONTROL") with a 28
V volt-
age supply for an external noise source. By switching the supply voltage for an exter-
nal noise source on or off in the firmware, you can enable or disable the device as
required.
External noise sources are useful when you are measuring power levels that fall below
the noise floor of the R&S
ZNL itself, for example when measuring the noise level of an
amplifier.
In this case, you can first connect an external noise source (whose noise power level is
known in advance) to the R&S
ZNL and measure the total noise power. From this
value you can determine the noise power of the R&S
ZNL. Then when you measure
the power level of the actual DUT, you can deduct the known noise level from the total
power to obtain the power level of the DUT.
Remote command:
5.4
Amplitude
Access
: "Overview" > "Amplitude"
Amplitude settings are identical to the Spectrum application, except for a new scaling
function for I/Q Vector and Real/Imag results (see
For background information on amplitude settings see the R&S
ZNL User Manual.
The "Auto Settings" are described in
Chapter 5.9, "Adjusting Settings Automatically"
5.4.1
Amplitude Settings
Access
: "Overview" > "Amplitude"
Amplitude