Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
142
User Manual 1178.5989.02 ─ 06
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
<Threshold>
When the reference level is adjusted automatically using the
on page 142 command, the internal attenuators are also adjusted. In order to
avoid frequent adaptation due to small changes in the input signal, you can define a
hysteresis. This setting defines an upper threshold the signal must exceed (compared
to the last measurement) before the reference level is adapted automatically.
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:TRIGger
<State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON
, for example).
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Manual operation:
See
" Adjusting the Center Frequency Automatically ( Auto Fre-
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
ZNL or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Manual operation:
See
" Setting the Reference Level Automatically ( Auto Level )"
Configuring I/Q Analyzer Measurements