![Rohde & Schwarz ZNL Series User Manual Download Page 167](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893167.webp)
Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
167
User Manual 1178.5989.02 ─ 06
Return values:
<Seconds>
Number of seconds that have passed since 01.01.1970 till the
frame start
<Nanoseconds>
Number of nanoseconds that have passed
in addition to the
<Seconds>
since 01.01.1970 till the frame start.
<Reserved>
The third value is reserved for future uses.
<Reserved>
The fourth value is reserved for future uses.
Example:
CALC:SGR:TST ON
Activates the time stamp.
CALC:SGR:TST:DATA? ALL
Returns the starting times of all frames sorted in a descending
order.
Usage:
Query only
Manual operation:
See
CALCulate<n>:SGRam:TSTamp[:STATe]
<State>
CALCulate<n>:SPECtrogram:TSTamp[:STATe]
<State>
This command activates and deactivates the time stamp.
If the time stamp is active, some commands do not address frames as numbers, but as
(relative) time values:
●
CALCulate<n>:DELTamarker<m>:SPECtrogram:FRAMe
●
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe
●
CALCulate<n>:SPECtrogram:FRAMe:SELect
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:SGR:TST ON
Activates the time stamp.
Manual operation:
See
9.7.2.2
Configuring the Color Map
DISPlay[:WINDow<n>]:SGRam:COLor:DEFault
................................................................ 168
DISPlay[:WINDow<n>]:SPECtrogram:COLor:DEFault
........................................................168
DISPlay[:WINDow<n>]:SGRam:COLor:LOWer
..................................................................168
DISPlay[:WINDow<n>]:SPECtrogram:COLor:LOWer
......................................................... 168
I/Q Analysis