Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
121
User Manual 1178.5989.02 ─ 06
Parameters:
<Volume>
Percentage of the maximum possible volume.
Range:
0 to 1
*RST:
0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
9.4.2
Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
................................................................................................ 121
...................................................................................122
9.4.2.1
Amplitude Settings
Useful commands for amplitude configuration described elsewhere:
●
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
........................................................... 121
........................................................................................... 121
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
........................................................122
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................... 122
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
<m>
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Manual operation:
See
" Reference Level = Marker Level "
CALCulate<n>:UNIT:POWer
<Unit>
UNIT<n>:POWer
<Unit>
This command selects the unit of the y-axis.
Configuring I/Q Analyzer Measurements