Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
105
User Manual 1178.5989.02 ─ 06
A measurement is started immediately with the default settings when the channel
setup is activated.
Different remote modes available
In remote control, two different modes for the I/Q Analyzer measurements are availa-
ble:
●
A quick mode for pure data acquisition
This mode is activated by default with the
command. The
evaluation functions are not available; however, performance is slightly improved.
●
A more sophisticated mode for acquisition and analysis.
This mode is activated when a new channel setup is opened for the I/Q Analyzer
application (
INST:CRE:NEW
/
INST:CRE:REPL
) or by an additional command (see
...................................................................................... 105
............................................................................................ 105
........................................................................................ 106
................................................................................. 108
INSTrument:CREate:DUPLicate
This command duplicates the currently selected channel setup, i.e creates a new chan-
nel setup of the same type and with the identical measurement settings. The name of
the new channel setup is the same as the copied channel setup, extended by a con-
secutive number (e.g. "IQAnalyzer" -> "IQAnalyzer 2").
The channel setup to be duplicated must be selected first using the
INST:SEL
com-
mand.
Usage:
Event
INSTrument:CREate[:NEW]
<ChannelType>, <ChannelName>
This command adds an additional measurement channel. You can configure up to 10
measurement channels at the same time (depending on available memory).
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
<ChannelName>
String containing the name of the channel.
Note that you can not assign an existing channel name to a new
channel; this will cause an error.
Activating I/Q Analyzer Measurements