Remote Control Commands
R&S
®
SMC100A
314
Operating Manual 1411.4060.32 ─ 04
:SYSTem:STARtup:COMPlete?
The command queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP
Response: 1
The startup of the instrument is completed.
Usage:
Query only
:SYSTem:TIME
<Time>
The command queries / sets the time for the instrument-internal clock.
The parameter is protected, in order to prevent accidental changes. It can be accessed
with protection level 1, see
Parameters:
<Time>
0...23,0...59,0...59
Example:
SYST:TIME?
Response:
12,0,0
it is precisely 12 pm.
SYSTem:VERSion?
The command queries the SCPI version with which the instrument complies.
Return values:
<Version>
string
Example:
SYST:VERS
queries the SCPI version.
Response: 1996
The instrument complies with the version from 1996.
Usage:
Query only
7.15 TEST Subsystem
The TEST system contains the commands for performing the routines as well as for direct
manipulation of the hardware assemblies (
:TEST:DIRect
).
The self tests return a "0" if the test is performed successfully, otherwise a value other
than "0" is returned. None of the commands of this system have an
*RST
value.
TEST Subsystem