Remote Control Commands
R&S
®
SMC100A
280
Operating Manual 1411.4060.32 ─ 04
Example:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Usage:
SCPI confirmed
[:SOURce<hw>]:POWer:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output level.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
The instrument operates at a constant level.
CW
and
FIXed
are
synonyms. The output level is set with
:SOURce:POWer
.
SWEep
The instrument operates in SWEep mode.
The level is set using the commands
SOURce:POWer:STARt;
STOP; MANual
.
*RST:
CW
Example:
POW:MODE SWEep
sets the SWEep mode. The settings under
SOURce:POW:STARt;
STOP; MANual
become effective.
Usage:
SCPI confirmed
[:SOURce<hw>]:POWer:POWer
<Power>
Sets the RF level of the RF output connector.
The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note:
The SCPI command
sets the level of the the "Level" display, i.e. the level containing offset.