![Rohde & Schwarz SMC100A Operating Manual Download Page 303](http://html.mh-extra.com/html/rohde-and-schwarz/smc100a/smc100a_operating-manual_1477935303.webp)
Remote Control Commands
R&S
®
SMC100A
286
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Source>
INTernal | EXTernal
INTernal
The internally generated rectangular signal is used for the pulse
modulation. The frequency of the internal signal can be set in the
SOURce:LFOutput subsystem.
EXTernal
The signal applied externally via the EXT MOD connector is used
for the pulse modulation.
*RST:
INTernal
Example:
PULM:SOUR INT
selects the internal modulation source.
Example:
PULM:STAT ON
activates the pulse modulation.
Usage:
SCPI confirmed
[:SOURce<hw>]:PULM:STATe
<State>
The command activates/deactivates the pulse modulation.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
PULM:STAT ON
activates pulse modulation with the modulation source selected
under
SOURce:PULM:SOURce
.
Usage:
SCPI confirmed
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity
<Polarity>
Selects the polarity of the Gate signal.
The signal is supplied via the PULSE EXT connector.
Parameters:
<Polarity>
NORMal | INVerted
*RST:
NORMal
Example:
PULM:TRIG:EXT:GAT:POL NORM
The pulse signal is generated while the gate signal is high.
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe
<Slope>
The command sets the polarity of the active slope of an applied trigger at the PULSE
EXT connector.
SOURce Subsystem