Error Messages
R&S
®
SMC100A
329
Operating Manual 1411.4060.32 ─ 04
The errors are assigned negative numbers. The error text being entered into the error/
event queue or being displayed is printed in bold face on the left together with the error
code. Below the error text, there is an explanation as to the respective error.
9.4 Device-Specific Error Messages
The following table contains all error messages specific for the instrument in alphabetical
order, as well as an explanation of the error situation. The positive error codes mark the
errors specific of the instrument.
The device-specific error messages set bit 3 in the ESR register.
The index provides a list of the error messages sorted according to their error codes.
Error
Code
Error
Description
Remedy
50
Extern reference out of
range or disconnected
External reference is selected but no
external signal is applied or the signal
is out of range.
●
Check the selected refer-
ence signal source (internal
or external) in the "Setup >
Reference Oscillator" dialog.
●
Change setting to 'internal' if
no appropriate external
source is available.
140
This modulation forces
other modulations off
A modulation has been switched on
which cannot be used at the same time
as an already active modulation. The
previous modulation has been
switched off.
Example: Enabling FM modulation
switches PM modulation off.
180
Adjustment failed
Adjustment could not be executed
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the device (see
ter 5.2.3.1, "Internal Adjust-
ments"
182
Adjustment data miss-
ing
Adjustment data are missing.
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the instrument.
183
Adjustment data invalid Adjustment data are invalid and must
be restored.
The adjustment data have to be
generated again by an internal or
external adjustment or to be loa-
ded into the instrument.
200
Cannot access hard-
ware
The data transmission to a module was
unsuccessful.
The module is not installed, not
properly installed or missing.
Device-Specific Error Messages