Remote Control Basics
R&S
®
SMC100A
173
Operating Manual 1411.4060.32 ─ 04
The instrument is switched to remote control as soon as it receives a command from
the controller.
While remote control is active, operation via the front panel or via mouse and key-
board is disabled and "REMOTE" is displayed in the status line.
The instrument remains in the remote state until it is reset to the manual state via the
instrument or via the remote control interface (see
chapter 6.2.2, "Returning to Man-
Tip:
Switching from manual operation to remote control and vice versa does not affect
the other instrument settings.
2. Although operation via front panel, mouse and keyboard is disabled, the dialog boxes
can still be opened, e.g. to verify settings, but buttons and setting fields are displayed
in gray and cannot be activated.
Use the command
SYST:KLOC ON
to disable the access to the dialogs.
3. To prevent unintentional return to manual operation, disable the LOCAL key of the
instrument using the
&LLO
chapter 6.1.3.3, "LAN Interface Mes-
The instrument switches to "REM-LLO" state and transition to manual mode is not
anymore possible via the remote control command
*GTL
.
Switching to manual mode is only possible via remote control then.
4. Enable the LOCAL key with the interface message
&NREN
.
6.2.2 Returning to Manual Operation
Before returning to manual control, command processing must be completed. Otherwise,
the instrument switches back to remote control immediately.
► To return to manual operation, use one of the following ways:
a) Press the LOCAL key on the front panel.
b) Select "Setup > Remote Control Channels" and press "Local".
c) While using the socket communication, terminate the remote control session.
d) Send the interface command
>L
via the remote control interface.
Tip:
Use the
&NREN
to enable the LOCAL key if the key is disabled.
6.2.3 Examples
This sections provides examples for setting up of remote control connection and starting
a remote control session over LAN and GPIB interfaces.
This section assumes basic knowledge of programming and operation of the controller.
A description of the interface commands can be obtained from the relevant manuals.
Starting a Remote Control Session