Remote Control Commands
R&S
®
SMC100A
261
Operating Manual 1411.4060.32 ─ 04
As with the "Frequency" value entered in the header, the
OFFSet
value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet
is set to 0. The value range for other
OFFset
values can be calculated using
the following formula:
300 kHz +
OFFSet
... RF
max
+
OFFSet
Parameters:
<Stop>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
500 MHz
Default unit: Hz
Example:
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
FREQ:STAR 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
<Increment>
The command sets the step width for the frequency setting if the frequency values
UP/DOWN
are used and variation mode
SOUR:FREQ:STEP:MODE USER
is selected. The
command is linked to "Variation Step" for manual control, i.e. the command also sets the
step width of the rotary knob for "Variation Active" on.
Parameters:
<Increment>
float
Range:
0Hz to RFmax - 100kHz
Increment: 0.01Hz
*RST:
1MHz
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
[:SOURce<hw>]:FREQuency:STEP:MODE
<Mode>
This command activates (
USER
) or deactivates (
DECimal
) the user-defined step width
used when varying the frequency value with the frequency values
UP/DOWN
. The com-
mand is linked to the command "Variation Active" for manual control, i.e. the command
also activates/deactivates the user-defined step width used when varying the frequency
value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
SOURce Subsystem