![Rohde & Schwarz SMC100A Operating Manual Download Page 236](http://html.mh-extra.com/html/rohde-and-schwarz/smc100a/smc100a_operating-manual_1477935236.webp)
Remote Control Commands
R&S
®
SMC100A
219
Operating Manual 1411.4060.32 ─ 04
Example:
HCOP:DEV:LANG BMP
selects output format *,bmp.
HCOP:FILE:AUTO:DIR '/usb/HCopy'
defines the destination directory '/usb/HCopy'.
HCOP:FILE:AUTO:PREF 'gen'
the file name starts with the prefix 'gen'. The usage of automatic
naming with prefix and date in the file name is preset
(…:STAT
ON
).
HCOP
triggers the generation of a hardcopy of the current trace.
HCOP:FILE:AUTO?
queries the path including the file name
Response:
/usb
/HCopy/gen101012008001.bmp
'
Usage:
Query only
:HCOPy:FILE[:NAME]:AUTO:DIRectory
<Directory>
This command defines the directory into which the hardcopy files will be stored if auto
naming is activated (
HCOP:FILE:AUTO:STAT ON
).The directory will be created if it does
not exist yet.
Parameters:
<Directory>
string
*RST:
.\HCopy
Example:
HCOP:FILE:AUTO:DIR '/usb/HCopy'
defines the destination directory '/usb/HCopy'
:HCOPy:FILE[:NAME]:AUTO:DIRectory:CLEar
This command deletes all files with extensions "bmp", "img", "png" and "xpm" in the
directory set for automatic naming.
Example:
HCOP:FILE:AUTO:DIR:CLE
deletes all image files with extensions "bmp", "img", "png" and
"xpm".
Usage:
Event
:HCOPy:FILE[:NAME]:AUTO:FILE?
This command queries the file name that what generated using the automatic naming
settings. By default the automatically generated file name is composed of:
<Prefix><YYYY><MM><DD><Number>.<Format>.
Each component can be deactivated/activated separately to individually design the file
name.
HCOPy Subsystem