Remote Control Commands
R&S
®
SMC100A
304
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Lterminator>
STANdard | EOI
EOI
The terminator must be sent together with the line message EOI
(End of Line). This setting is recommended for binary block trans-
missions where a character could coincidentally have the value LF
(Line Feed) but is not intended as the terminator. This setting must
be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of
whether it is sent with or without EOI.
*RST:
STANdard
Example:
SYST:COMM:GPIB:LTER EOI
only a character which is sent simultaneously with the line mes-
sage EOI is accepted as the terminator.
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
The command sets the GPIB address.
Parameters:
<Address>
float
Range:
1 to 30
*RST:
28
Example:
SYST:COMM:GPIB:ADDR 14
sets GPIB address.
Usage:
SCPI confirmed
:SYSTem:COMMunicate:NETWork[:COMMon]:DOMain
<Domain>
The command enters the domain of the network.
Parameters:
<Domain>
string
Example:
SYST:COMM:NETW:COMM:DOM ABC.DE
enters the domain of the network.
:SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
<Hostname>
The command enters the individual computer name of the R&S Signal Generator. The
entry is only activated after sending command
SYST:COMM:NETW:COMM:SET
.
Note:
The hostname can only be changed after deactivating protection level 1 (command
SYSTem:PROTect<n>:STATe OFF
, password)
Parameters:
<Hostname>
string
SYSTem Subsystem