Hardware Interfaces
R&S
®
SMC100A
332
Operating Manual 1411.4060.32 ─ 04
SRQ
(Service Request): active LOW enables the connected device to send a service
request to the controller.
REN
(Remote Enable): active LOW permits switchover to remote control.
EOI
(End or Identify): has two functions in connection with ATN:
–
ATN=HIGH active LOW marks the end of data transmission.
–
ATN=LOW active LOW triggers a parallel poll.
●
Handshake bus with three lines:
DAV
(Data Valid): active LOW signals a valid data byte on the data bus.
NRFD
(Not Ready For Data): active LOW signals that one of the connected devices
is not ready for data transfer.
NDAC
(Not Data Accepted): active LOW signals that the instrument connected is
accepting the data on the data bus.
Interface Functions
Instruments which can be controlled via GPIB bus can be equipped with different interface
functions. The interface function for the R&S SMC are listed in the following table.
Table 1-1: GPIB bus interface functions
Control character
Interface function
SH1
Handshake source function (source handshake), full capability
AH1
Handshake sink function (acceptor handshake), full capability
L4
Listener function, full capability, de-addressed by MTA.
T6
Talker function, full capability, ability to respond to serial poll, deaddressed
by MLA
SR1
Service request function (Service Request), full capability
PP1
Parallel poll function, full capability
RL1
Remote/Local switch over function, full capability
DC1
Reset function (Device Clear), full capability
DT1
Trigger function (Device Trigger), full capability
GPIB Bus Interface