Remote Control Commands
R&S
®
SMC100A
214
Operating Manual 1411.4060.32 ─ 04
Return values:
<OTIMe>
float
Example:
DIAG:INFO:OTIM
queries the operation hours.
Response: 100023
The instrument was operated for 100023 hours up to now.
Usage:
Query only
:DIAGnostic:INFO:POCount?
The command queries the number of power-on events.
Return values:
<Pocount>
float
Example:
DIAG:INFO:POC
queries the number of power on events.
Response: 123
The instrument was switched on for 123 times up to now.
Usage:
Query only
7.5 DISPlay Subsystem
The DISPlay system contains the commands to set the power-save mode of the instru-
ment.
:DISPlay:PSAVe:HOLDoff
<Holdoff>
This command sets the wait time for the screen-save mode of the display. The available
value range is 1 to 60 minutes, the resolution 1 minute. The entry is dimensionless.
Parameters:
<Holdoff>
float
Range:
1 to 60
*RST:
1
Default unit: minute
Example:
DISP:PSAV:HOLD 10
sets the wait for the screen saver mode to 10 minutes.
:DISPlay:PSAVe[:STATe]
<State>
This command activates/deactivates the screen-save mode of the display. With the
screen-save mode activated the display including backlight is completely switched off
after the elapse of the wait time (see command
DISPlay:PSAVe:HOLDoff
) when no
entries via front panel, external mouse or external keyboard are made.
DISPlay Subsystem