Error Messages
R&S
®
SMC100A
327
Operating Manual 1411.4060.32 ─ 04
REM-LLO
The instrument is remote (
REM
ote) controlled. The LOCAL key is disabled by remote
control with the command
LLO
(
L
ocal
L
ock
O
ut).
The keys on the front panel are usable, but all parameters are in read only mode.
The instrument can be switched from remote control to manual operation by means of
remote control only (e.g. with the Visual Basic command
CALL IBLOC
(generator%)
or by
&NREN
).
LOC-LLO
For operating directly the instrument is placed from remote control to manual operation
(Local State). The LOCAL key was disabled by remote control with the command
LLO
(
L
ocal
L
ock
O
ut).
With the next activating of the remote control mode, the instrument cannot be switched
to manual operation by the operator. The status information changes to "REM-LLO".
The instrument can be switched to manual operation by means of remote control only
(e.g. with the Visual Basic command
CALL IBLOC (generator%)
).
9.1.3 Status information displayed in the Info line
This chapter gives an overview of the status messages displayed in the Info line.
RFSweep / LevelSweep / LFSweep
The indicated sweep is enabled.
ALC On / Auto / S&H
The status of the automatic level control is indicated:
●
ON
automatic level control permanently on
●
Auto
automatic level control is automatically adapted to the operating states
●
S&H
automatic level control off, recalibration of the level whenever the level or frequency
is set (sample and hold mode)
AttFixed
Attenuator fixed mode is active.
The uninterrupted level settings are made in a fixed range without attenuator switching.
The variation range is set automatically when this mode is activated. The range is dis-
played under "Attenuator Fixed Range" in the "Level" dialog.
UCorr
User Correction is active.
The level is corrected by the given values in the selected user correction list. Correction
is performed by the user-defined list values being added to the output level for the
respective RF frequency. With frequencies which are not contained in the list, the level
correction is determined by interpolation of the closest correction values.
Status Information