Remote Control Commands
R&S
®
SMC100A
282
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Increment>
float
Range:
0 dB to 100 dB
Increment: 0.01 dB
*RST:
1 dB
Default unit: dB
Example:
SOURce:POWer:STEP:INCRement 2
or
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
[:SOURce<hw>]:POWer:STEP:MODE
<Mode>
This command activates (
USER
) or deactivates (
DECimal
) the user-defined step width
used when varying the level value with the level values
UP/DOWN
. The command is linked
to setting "Variation Active" for manual control, i.e. the command also activates/deacti-
vates the user-defined step width used when varying the level value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
Example:
POW:STEP 2
sets the step width for the level setting to 2 dB.
POW:STEP:MODE USER
actives this step width for level variation with the rotary knob (man-
ual control) and with level values
UP/DOWN
(remote control).
[:SOURce<hw>]:POWer:STOP
<Stop>
The command sets the stop level in Sweep mode.
As with the
Level
value entered in the
RF Level
menu, the
OFFSet
value is also taken
into consideration with this command.
The specified value range is therefore only effective if
:SOURce:POWer:OFFSet
is set
to 0. The value range for other
OFFset
values can be calculated using the following
formula:
Minimum level +
OFFSet
... Maximum level +
OFFSet
Parameters:
<Stop>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model.
The values are given in the data sheet.
Increment: 0.01 dB
*RST:
-10 dBm
Default unit: dBm
SOURce Subsystem