![Rohde & Schwarz SMC100A Operating Manual Download Page 223](http://html.mh-extra.com/html/rohde-and-schwarz/smc100a/smc100a_operating-manual_1477935223.webp)
Remote Control Commands
R&S
®
SMC100A
206
Operating Manual 1411.4060.32 ─ 04
Sets the status byte (STB), the standard event register (ESR) and the
EVENt
part of the
QUEStionable
and the
OPERation
registers to zero. The command does not alter the
mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE
<Value>
Event Status Enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*ESR?
Event Status Read
Returns the contents of the event status register in decimal form and subsequently sets
the register to zero.
Return values:
<Contents>
Range:
0 to 255
Usage:
Query only
*IDN?
IDeNtification: returns the instrument identification.
Return values:
<ID>
"Rohde&Schwarz,<device type>,<serial number>,<firmware ver-
sion>"
Example:
Rohde&Schwarz,ZVA8-4Port,12345,0.10.1.23
Usage:
Query only
*IST?
Individual STatus query
Returns the contents of the IST flag in decimal form. The IST flag is the status bit which
is sent during a parallel poll.
Return values:
<ISTflag>
0 | 1
Usage:
Query only
Common Commands