Instrument Function
R&S
®
SMC100A
86
Operating Manual 1411.4060.32 ─ 04
The dialog provides an access to the network settings, like settings about the general
network environment and specific identification of the computer in the network, as well
as an indication whether the instrument is connected to the network or not.
Network Status
Indicates whether the instrument is connected to the network or not.
SCPI command:
n.a.
Hostname
Enters the individual computer name of the R&S Signal Generator.
Note:
The computer name can only be changed after protection level 1 is deactivated
chapter 5.2.3.12, "Protection"
A predefined name is indicated and can be used for network connections.
SCPI command:
Workgroup
Enters the individual windows workgroup name of the R&S Signal Generator. This
parameter is necessary in case the instrument is integrated in a windows network.
Note:
The workgroup can only be changed after protection level 1 is deactivated (see
chapter 5.2.3.12, "Protection"
SCPI command:
Address Mode
Selects if the IP address is assigned automatically or manually.
"Auto (DHCP)" The IP address is assigned automatically.
The network used must support automatic assignment of IP address
via DHCP or APIPA (Zeroconf) in order to use this function.
"Static"
The IP address is assigned manually.
SCPI command:
IP Address
Displays the IP address. To enter the IP address manually, select "Static" Address Mode.
SCPI command:
Subnet Mask
Displays the Subnet mask. To enter the Subnet mask manually, select "Static" Address
Mode.
This number is used together with the IP address to identify the network segment the
instrument is in.
SCPI command:
General Instrument Settings