Instrument Function
R&S
®
SMC100A
143
Operating Manual 1411.4060.32 ─ 04
"Extern Single" Sets a single sweep cycle. The sweep is triggered by an external trigger
signal.
If a different sweep mode was activated prior to the "Extern Single"
mode, the current sweep is stopped. The "Extern Single" sweep always
starts at the start level.
Refer to the description of the rear panel for information about the con-
nectors for external trigger signal input (see
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE
(External trigger)
"Extern Step"
Sets a step-by-step sweep cycle. Each sweep step is triggered by an
external trigger signal (trigger source as described under "Extern Sin-
gle"). The step width corresponds to the step width of the rotary knob.
If a different sweep mode was activated prior to the "Extern Step" mode,
the current sweep is stopped. The "Extern Step" sweep always starts
at the start level.
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE STEP
SOUR:SWE:POW:STEP 0.5
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE
(External trigger)
RF Signal - RF Block