Remote Control Commands
R&S
®
SMC100A
266
Operating Manual 1411.4060.32 ─ 04
[:SOURce]:LFOutput[:STATe]
<State>
The command activates/deactivates the LF output.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
LFO ON
activates the LF output. The settings under
LFO:FREQ
and
LFO:SWE
become effective.
[:SOURce]:LFOutput:SWEep[:FREQuency]:DWELl
<Dwell>
The command sets the dwell time for each frequency step of the sweep.
Tip:
It is recommended to switch off the "GUI Update" for optimum sweep performance
especially with short dwell times (
).
Parameters:
<Dwell>
float
Range:
10.0E-3 to 100
Increment: 100.0E-6
*RST:
10.0E-3
Default unit: s
Example:
LFO:SWE:DWEL 20 ms
sets a dwell time of 20 ms.
[:SOURce]:LFOutput:SWEep[:FREQuency]:EXECute
The command immediately starts an LF sweep.
The command is only valid for sweep mode Single
(
SOURce:LFOutput:SWEep:FREQuency:MODE SINGle
).
The command corresponds to the manual-control
Execute Single Sweep - LF Sweep
Example:
LFO:SWE:MODE SING
sets the single cycle mode of the LF sweep.
LFO:SWE:EXEC
starts one cycle of the LF sweep.
Usage:
Event
[:SOURce]:LFOutput:SWEep[:FREQuency]:MODE
<Mode>
The command sets the cycle mode of the LF sweep.
The assignment of the GPIB commands to the sweep modes is given in the description
of the sweep menus.