Remote Control Commands
R&S
®
SMC100A
313
Operating Manual 1411.4060.32 ─ 04
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:ORES "Test Option"
defines the OPT string 'test option'
*OPT?
Response: 'test option'
:SYSTem:PROTect<ch>[:STATe]
<State>,<password>
The command activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
There are several protection levels which disable specific service
functions (authorized personnel of R&S Service Departments
only).
Parameters:
<State>
0 | 1 | OFF | ON
<password>
The respective functions are disabled when the protection level is
activated. No password is required for activation. A password must
be entered to deactivate the protection level. The password for the
first level is 123456. This protection level can be used to lock-out
internal adjustments.
Example:
SYST:PROT1 ON
activates protection level 1. Internal adjustments are only possible
after deactivating the lock-out.
SYST:PROT1 OFF, 123456
deactivates protection level 1. Internal adjustments are enabled
again.
Usage:
SCPI confirmed
SYSTem:SERRor?
This command returns a list of all errors existing at the time when the query is started.
This list corresponds to the display on the info page under manual control.
Return values:
<Serror>
string
Example:
SYST:SERR
queries all errors existing in the error queue.
Example:
Response: -221, 'Settings conflict', 153, 'Input
voltage out of range'
The two returned errors have occurred since the error queue was
last queried.
Usage:
Query only
SYSTem Subsystem