NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
69
User Manual 1178.5689.02 ─ 06
Out of Sync
The "Out of Sync" timer specifies the time after which the instrument, having waited for
a signal from the connected UE, releases the connection. The timer is started when an
uplink grant message is sent to the UE.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:TOUT:OOSYnc
T3324
Timer "T3324" defines how long the UE stays in idle mode after a transition from RRC
connected to RRC idle. When the timer expires, the UE enters the power-saving mode
(PSM).
The T3324 value is only signaled to the UE if the UE requests PSM and PSM is
allowed, see
If the checkbox is off, the value requested by the UE is returned. If the checkbox is on,
the configured value is returned.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:TOUT:T3324
T3412
Timer "T3412" controls the initiation of a periodic tracking area update by the UE.
If the timer is disabled, no periodic tracking area update is required.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:TOUT:T3412
T3412 Extended
The extended T3412 setting allows much longer timeout values than the T3412 setting.
If the UE supports extended values for T3412, it uses this setting instead of the T3412
setting.
Configure the timer in seconds. The setting is displayed for information in days, hours,
minutes and seconds.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:TOUT:TEXT
2.3.10.7
Time
Access: "Config" hotkey > "Network" > "Time" section.
The "Time" section allows you to send configurable date and time information to the
UE. Thus you can update the date and time displayed by the mobile. In a real network,
this service is typically used to send the current local time to the UE.
GUI Reference