NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
312
User Manual 1178.5689.02 ─ 06
<2_OutOfTol>
Out of tolerance result, i.e. percentage of measurement inter-
vals of the statistic count for modulation measurements exceed-
ing the specified modulation limits.
Default unit: %
<3_EVM_RMS>
Error vector magnitude RMS value
Default unit: %
<4_EVMpeak>
Error vector magnitude peak value
Default unit: %
<5_MagErrorRMS>
Magnitude error RMS value
Default unit: %
<6_MagErrPeak>
Magnitude error peak value
Default unit: %
<7_PhErrorRMS>
Phase error RMS value
Default unit: deg
<8_PhErrorPeak>
Phase error peak value
Default unit: deg
<9_IQOffset>
I/Q origin offset
Default unit: dBc
<10_FreqError>
Carrier frequency error
Default unit: Hz
<11_TimingError>
Transmit time error
Default unit: Ts
<12_TXpowerMin>
Minimum user equipment power
Default unit: dBm
<13_TXpowerMax>
Maximum user equipment power
Default unit: dBm
<14_PeakPowerMin> Minimum user equipment peak power
Default unit: dBm
<15_PeakPowerMax>Maximum user equipment peak power
Default unit: dBm
<16_SCPowerMin>
Minimum SC power
Default unit: dBm
<17_SCPowerMax> Maximum SC power
Default unit: dBm
<18_EVM_DMRS>
Error vector magnitude DMRS value
Default unit: %
Command Reference
深圳德标仪器