NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
158
User Manual 1178.5689.02 ─ 06
OFF
The configured timer value is ignored and the user-requested
value is returned to the UE after a PSM request.
Range:
2 s to 11160 s
Increment: 2 s (2 s to 62 s), 60 s (120 s to 1860 s), 360 s (2160
s to 11160 s)
*RST:
OFF (2 s)
Example:
See
Firmware/Software:
V3.7.30
Manual operation:
See
CONFigure:NIOT:SIGN<i>:CELL:TOUT:T3412
<Value>
Configures timer T3412 for periodic tracking area updates.
The information element supports the values 1 to 31 combined with the units 2 sec-
onds, 1 minute and 6 minutes.
This command configures the timer value in seconds. So there are three subranges
with different increments.
Parameters:
<Value>
The parameter configures a checkbox (ON | OFF) or a timer
value.
ON
Enables the timer and transmission of the timer value.
Setting a timer value sets also ON.
OFF
Disables the timer and transmission of the timer value.
Range:
2 s to 11160 s
Increment: 2 s (2 s to 62 s), 60 s (120 s to 1860 s), 360 s (2160
s to 11160 s)
*RST:
OFF (2 s)
Example:
See
Firmware/Software:
V3.7.30
Manual operation:
See
CONFigure:NIOT:SIGN<i>:CELL:TOUT:TEXT
<value>
Configures an extended value for timer T3412.
The information element supports the values 1 to 31 combined with the units 2 s, 30 s,
1 min, 10 min, 1 h, 10 h and 320 h.
This command configures the timer value in seconds. So there are subranges with dif-
ferent increments.
Command Reference