NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
67
User Manual 1178.5689.02 ─ 06
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:CID:EUTRan
2.3.10.4
Security
Access: "Config" hotkey > "Network" > "Security" section.
The "Security Settings" configure parameters for the authentication procedure and for
integrity protection.
Figure 2-38: Security settings
Authentication
Enables or disables authentication, to be performed during the attach procedure.
Authentication requires a test USIM. An appropriate 3GPP USIM can be obtained from
Rohde & Schwarz (R&S CMW-Z06, stock no. 1209.2010.02).
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:SECurity:AUTHenticat
Security
Enables or disables security. With enabled security, the UE uses integrity protection for
signaling. This setting is only relevant if authentication is enabled.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:SECurity:SECurity
Integrity Algorithm
Selects an algorithm for integrity protection.
NULL means that integrity protection is disabled. Use this setting for UEs which do not
support any integrity algorithm.
Remote command:
CONFigure:NIOT:SIGN<i>:CELL:SECurity:IALGorithm
Secret Key
The secret key K is used for the authentication procedure (including a possible integrity
check). The value is entered as 32-digit hexadecimal number.
GUI Reference
深圳德标仪器
135-1095-0799