Refer also to the following commands:
●
CONFigure:NIOT:SIGN<i>:RFSettings:ENPMode
●
CONFigure:NIOT:SIGN<i>:RFSettings:ENPower
●
CONFigure:NIOT:SIGN<i>:RFSettings:EATTenuation:INPut
Parameters:
<UserMargin>
Range:
0 dB to (55 dB + external attenuation - expected
nominal power)
*RST:
12 dB
Default unit: dB
Example:
See
Firmware/Software:
V3.7.20
Manual operation:
See
"Exp. Nominal Power..., Margin"
CONFigure:NIOT:SIGN<i>:RFSettings:MLOFfset
<MixLevOffset>
Varies the input level of the mixer in the analyzer path.
Parameters:
<MixLevOffset>
Range:
-10 dB to 10 dB
*RST:
0 dB
Default unit: dB
Example:
See
Firmware/Software:
V3.7.20
Manual operation:
See
2.5.6.4
User-Defined Band Configuration
The following commands define user-specific operating band.
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:CHANnel:UL:MAXimum?
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:FREQuency:DL:MAXimum?
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:FREQuency:UL:MAXimum?
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:FREQuency:UL:MINimum?
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:SET
...................................................... 132
CONFigure:NIOT:SIGN<i>:RFSettings:UDEFined:CHANnel:UL:MAXimum?
Queries the maximum channel number in uplink for user-defined operating band.
Return values:
<Channel>
Example:
See
Configuring Operating Bands and Channels