NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
89
User Manual 1178.5689.02 ─ 06
The "Debug" node configures the log file generation.
Do not enable the debug mode during regular operation. It impacts the test perform-
ance, timing, system drive usage, and so on.
Figure 2-55: Debug configuration
The settings are not affected by a reset or preset. They are not stored in save files. A
reboot restores the default settings.
.........................................................................................89
................................................................................................. 89
Signaling Unit Shutdown During Signal Off
...................................................................89
................................................................................ 90
Write Logs Now
Writes the NB-IoT log files to the system drive.
Cell Restart with Logs Reset
Switches the cell signal off and on again with empty logs.
Enable Debug Settings
Enables the configuration of the "Debug Settings".
Signaling Unit Shutdown During Signal Off
The complete log files are written if the signaling unit is shut down. The signaling unit is
shut down if the instrument is shut down or if the CMW software is terminated.
With enabled checkbox, the signaling unit is also shut down if the cell signal is
switched off. So switching off the cell delivers log files.
Complete Logs
With disabled checkbox, only a reduced set of log files is written if the NB-IoT signaling
instance is replaced by another signaling instance due to a resource conflict. This
behavior saves time and system drive space.
With enabled checkbox, a complete set of log files is written if the NB-IoT signaling
instance is replaced by another signaling instance.
GUI Reference