2. Important Safety Information
2-4 Safety information
27 / 168
<Components that can be connected (GD-3)>
The following components can be connected to the remote sensor head (GD-3):
Connector
1
Connector
2
Connector
3
Sensor assembly
×
×
〇
EC barrier
×
×
〇
Approved cable gland
〇
〇
×
M25 plug
〇
〇
〇
Adapter
(M25
⇔
1/2NPT)
〇
〇
〇
Adapter
(M25
⇔
3/4NPT)
〇
〇
〇
Adapter
(M25
⇔
M20)
〇
〇
〇
〇
: Can be connected
×
: Cannot be connected
NOTE
Connect only components that are compatible with the individual connectors on the remote sensor head
(GD-3).
Tighten M25 plugs and adapters to a torque of at least 30 Nm.
Secure a minimum thread engagement of five threads.