7. Maintenance
7-2 Maintenance mode
111 / 168
4 Once the operation is complete, hold down the MENU/ESC key (for approximately three seconds).
The product switches to detection mode.
WARNING
Once the operation is complete, be sure to hold down the MENU/ESC key to return to detection mode.
The product will not detect gases while in adjustment/setting state. If left in maintenance mode, the
product will take 10 hours to return to detection mode. Detection will not be available during this time,
posing hazards.
NOTE
If the password setting is enabled, the password must be entered to access maintenance mode.
The password is set in advance by Riken Keiki.
If you are unsure, contact Riken Keiki.
<Switching from maintenance mode to user mode>
1 Press the ▲ or ▼ key in maintenance mode several times
to select [2-9], then press the SET key.
2 Press the SET key.
The product switches to user mode.