2. Important Safety Information
2-4 Safety information
23 / 168
CAUTION
Install the product in accordance with the required installation conditions.
Do not disassemble, modify, or alter the product.
Contact Riken Keiki to request readjustment or parts replacement including calibration.
Use cable lugs for grounding wire and bonding conductor connections. Make sure the bonding
conductors are not loose or twisted.
Do not open the cover in locations where explosive gas may be present.
The maximum pressure inside the gas channels and sensor is 10.13 kPa (gauge pressure).
Be sure to connect the grounding terminal before turning on the power to the product.
Make sure the grounding is Class D equivalent (grounding resistance less than 100 Ω).
Use crimped terminals to connect the grounding wire to the grounding terminal inside the unit. Use a
cable with a cross section area of at least 4 mm
2
to connect to the external grounding terminal.
Use approved cable glands when installing.
If the product is installed in a location where the ambient temperature exceeds 65 °C, use cables
designed for use at temperatures at least 5 °C above the maximum ambient temperature.
Do not subject the window in the cover to strong forces or impact. Doing so may result in damage and
impaired explosion-proof performance. (The low risk of mechanical hazards has been confirmed in
testing [2 joules].)
Do not use more than one adapter per connector.
CAUTION
With suction type
Use with a flow rate of at least 0.4 L/min and not exceeding 1.5 L/min.
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...