11. Appendix
11-2 Terminology
166 / 168
11-2 Terminology
Term
Definition
Initialization
This function prevents alarms from being issued for a while after the power is
turned on.
This prevents alarms from being issued, as the output from the detector
fluctuates for a while after the power is turned on.
Full scale
The maximum value of the detection range
Calibration
This refers to the process of checking the product reading against a calibration
gas.
The process includes adjusting the product reading to match the concentration
of the calibration gas.
Zero suppression
This function is used to hide (suppress) reading fluctuations below a set lower
limit to minimize sensor output fluctuations not caused by gas, such as
temperature and humidity fluctuations and noise effects.
Alarm delay time
This function temporarily suspends immediate alarm activation to prevent false
alarms caused by external noise, even if the reading reaches an alarm setpoint.
Inhibit
This function pauses gas detection—for example, when the product is
undergoing maintenance. It is also sometimes referred to as point skipping.
%
Unit used to express the percentage of a target gas contained in a given
volume
ppm
Unit used to express the parts per million of a target gas contained in a given
volume
%LEL
Unit for combustible gases in which the lower explosive limit (LEL) of the target
gas is set to 100 % (Example: For methane, 5 % concentration by volume
corresponds to 100 %LEL.)
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...