4. Installation
4-5 Remote sensor head (GD-3) installation
55 / 168
2 Insert the remote sensor head (GD-3) into the duct
mounting kit and secure with four M5 bolts.
Connect a tube to the sensor on the remote sensor head
(GD-3) to enable calibration. The tube should protrude from
the top of the duct mounting kit.
3 Insert the duct mounting kit into the opening in the duct.
Secure in place with six M6 bolts.
0
NOTE
When mounting the remote sensor head (GD-3) inside a duct,
make sure the sensor is not in contact with the duct inner wall.
Duct mounting kit
Duct
300 or m
ore
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...