5. Usage Instructions
5-6 Performing zero calibration
87 / 168
NOTE
If zero calibration is unsuccessful because the zero point ([20.9] (vol%) in the case of oxygen) fluctuates
significantly around zero, [FAIL] is displayed before the display returns to [1-1]. If this occurs, zero
calibration was not performed.
[1-1] corresponds to span adjustment for the oxygen deficiency alarm specifications. Introduce clean air
to adjust to [20.9] (vol%). For information on span adjustment, refer to ‘7-4-3 Performing span
adjustment’.