10. Product Specifications
10-3 Detection target gases for individual detection principles
160 / 168
10-3-5 Electrochemical type
Sensor model
Gas name
Gas
symbol
F.S.
1 digit
First
Alarm
setpoint
Second
Alarm
setpoint
Calibration
gas
Operating
temperature
range
Operating
humidity
range
Years of
warranty
ESF-A24R
Hydrogen
sulfide
H
2
S
100 ppm
1 ppm
20 ppm
40 ppm
H
2
S
−40 °C - 70 °C
*2
20 %RH -
90 %RH (no
condensation)
3 years
ESF-A24RH
*3
Hydrogen
sulfide
H
2
S
30 ppm
0.2 ppm
10 ppm
20 ppm
H
2
S
−40 °C - 70 °C
*1
40 %RH -
95 %RH (no
condensation)
3 years
*1: Allow the product to fully adjust to the usage conditions before adjusting if below −25 °C. Can be used for a total of
one week between 40 °C and 70 °C.
*2: Allow the product to fully adjust to the usage conditions before adjusting if below −25 °C. Can be used for a total of
one week between 55 °C and 70 °C.
NOTE
Perform maintenance if a warning appears while the life assessment or self-diagnostic function is active.
Use the ESF sensor adapter when calibrating. Calibration accuracy may be reduced without the ESF
sensor adapter.
Use a flow rate of 0.5 ± 0.1 L/min when calibrating. Calibration accuracy may be reduced if calibration is
performed using a different flow rate.
When using a diffusion type detector, gas sensitivity will be affected by ambient wind speeds.
The number of years of warranty for the ESF sensor depends on the sensor model. Refer to the above
table.
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...